Specifications and technique
SPECTROSCOPIC PERFORMANCE (typical, at 25 °Celsius with InAs detector) Spectral Range 3,700 to 14,885 cm-1 Signal to Noise (60 s, 16 cm-1, at peak response) > 900 000: 1
Frequency Repeatability (@ 7300 cm-1) < 0.006 cm-1
Frequency Accuracy (@ 7300 cm-1) < 0.01 cm-1
Absorbance Reproducibility (toluene) < 0.002 AU
OPTICAL BENCH
Beamsplitter Material ZnSe (non-hygroscopic)
Interferometer Mechanism (patent pending) High-throughput double pivot Michelson, fully jacketed
Source (with electronic stabilization) Quartz Halogen, 10-year expected life
Solid-state Laser Maintenance Free
DATA COMMUNICATIONS
Hardware Port Ethernet, 10/100 Mbps
INSTRUMENT ENCLOSURE
Material Rugged Design, all-metal frame with integral handles
Size 435 mm (W) x 280 mm (D) x 372 mm (H), Weight 24 kg
ENVIRONMENTAL
Universal Power Supply 120-240 VAC, 50/60 Hz, 65 W
Operating Temperature 10 °C to 35 °C
Operating Relative Humidity 5 % to 80 %, non-condensing
Regulatory Certification and Compliance TUV, CE and RoHS
Technique
 The Quant Analyser is powered by the latest ABB Bomem FT-NIR technology. FT-NIR is short for Fourier Transform Near Infrared and measures the entire spectrum of the sample instantly. This technique generates a large amount of high-quality spectral data, which makes it possible to accurately determine various components such as fat, moisture, protein, lactose etc.
|